Improved Tools for Point-of-Care Nucleic Acid Amplification Testing

Author: Jue, Erik Bradley

Year: 2020

Degree: Dissertation (Ph.D.)

Advisor: Ismagilov, Rustem F.

Committee Members: Murray, Richard M.; Ismagilov, Rustem F.; Shapiro, Mikhail G.; Yang, Changhuei

Option: Bioengineering

DOI: 10.7907/d6mf-5081

Abstract

There is a critical need for improved diagnostic tools to detect infectious diseases, especially in low-resource regions. A sample-to-answer point-of-care nucleic acid amplification test (NAAT) would be incredibly valuable for many different applications (e.g. COVID-19, Chlamydia/Gonorrhoeae, Influenza, Ebola, Zika/Chikungunya/Dengue, etc.). However, sample preparation (purification of pure nucleic acids) is a challenging bottleneck. In Chapter 2, commercial NA extraction methods were studied and improved. In Chapter 3, commercial stocks of SARS-CoV-2 RNA used in FDA emergency-use authorizations were found to be inaccurate and were independently quantified using reverse transcription digital PCR. In Chapter 4, a 3D printed meter-mix device was developed for initial processing prior to the sample preparation device. In Chapter 5, a 3D printed sample-to-device interface was prototyped to facilitate loading multi-volume SlipChip devices with purified template mixed with LAMP reactants. In Chapters 6-7, advancements were made for image processing of commercial chips to study digital LAMP reactions. In Chapter 8, additional tools were developed towards sample-to-answer point-of-care NAAT including a sample preparation module, amplification module, cell-phone readout, and automated base station.